Nitrate LR

• Replace the cap and immediately shake vigorously up

and down for exactly 10 seconds. Continue to mix by
inverting the cuvette gently for 50 seconds, while
taking care not to induce air bubbles. Powder will not
completely dissolve. Time and way of shaking could
sensitively affect the measurement.

• Reinsert the cuvette into the instrument, taking care

not to shake it.

• Press Timer and the display will show the countdown

prior to the measurement or, alternatively, wait for
4 minutes and 30 seconds and press Read. When
the timer ends the meter will perform the reading.
The instrument displays the results in mg/L of
nitrate-nitrogen.

• Press or to access the second level functions.
• Press the Chem Frm key to convert the result in mg/L of nitrate (NO

3

¯).

• Press or to return to the measurement screen.

INTERFERENCES

Interference may be caused by:
Ammonia and amines, as urea and primary aliphatic amines
Chloride above 100 ppm

Chlorine above 2 ppm
Copper

Iron(III)
Strong oxidizing and reducing substances
Sulfide must be absent