• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.

• Remove the cuvette and add the content of one packet

of HI 93750B-0 reagent. Replace the cap and gently
mix for one minute by slowly turning the cuvette upside
down.

• Reinsert the cuvette into the instrument.

• Press Timer and the display will show the countdown prior to

the measurement or, alternatively, wait for 2 minutes and
press Read. When the timer ends the meter will perform the
reading. The instrument displays the results in mg/L (ppm)
of potassium (K).

• Press or to access the second level functions.
• Press the Chem Frm key to convert the result in mg/L of potassium oxide (K

2

O).

• Press or to return to the measurement screen.

INTERFERENCES

Interferences may be caused by:

Ammonium above 10 ppm

Calcium above 10000 ppm as CaCO

3

Chloride above 12000 ppm

Magnesium above 8000 ppm as CaCO

3

Sodium above 8000 ppm

Potassium LR