36

Phosphate LR

PHOSPHATE LOW RANGE

SPECIFICATIONS

Range

0.00 to 2.50 mg/L

Resolution

0.01 mg/L

Accuracy

±0.04 mg/L ±4% of reading at 25 °C

Typical EMC Dev. ±0.01 mg/L
Light Source

Tungsten lamp with narrow band interference filter @ 610 nm

Method

Adaptation of the Ascorbic Acid method. The reaction between phosphate and the
reagent causes a blue tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93713-0

Powder reagent

1 packet

REAGENT SETS

HI 93713-01 Reagents for 100 tests
HI 93713-03 Reagents for 300 tests
For other accessories see page 49.

MEASUREMENT PROCEDURE

• Select the

Phosphate LR method using the procedure

described in the

Method Selection section (see page 12).

• Rinse, cap and shake the cuvette several times with unreacted

sample. Fill the cuvette with 10 mL of sample up to the
mark and replace the cap.

• Place the cuvette into the holder and close the lid.

• Press the Zero key. The display will show “-0.0-” when the meter is zeroed and ready for measurement.

• Remove the cuvette and add the content of one packet of

HI 93713-0 reagent. Replace the cap and shake gently (for
about 2 minutes) until the powder is completely dissolved.

10 mL

Advertising