Hanna Instruments HI 83210-01 User Manual

Page 38

Advertising
background image

38

Silica

SILICA

SPECIFICATIONS

Range

0.00 to 2.00 mg/L

Resolution

0.01 mg/L

Accuracy

±0.03 mg/L ±3% of reading reading at 25 °C

Typical EMC

±0.01 mg/L

Deviation

Light Source

Tungsten lamp with narrow band interference filter @ 610 nm

Method

Adaptation of the

ASTM Manual of Water and Environmental Technology, D859,

Heteropoly Blue method. The reaction between silica and reagents causes a blue tint in

the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93705A-0

Molybdate

6 drops

HI 93705B-0

Citric acid

1 packet

HI 93705C-0

Amino acid

1 packet

REAGENT SETS

HI 93705-01 Reagents for 100 tests
HI 93705-03 Reagents for 300 tests
For other accessories see page 49.

MEASUREMENT PROCEDURE

• Select the

Silica method using the procedure described in the

Method Selection section (see page 12).

• Fill the cuvette with 10 mL of unreacted sample (up to the

mark).

• Add 6 drops of HI 93705A-0 Molybdate reagent. Replace the

cap and swirl the solution.

• Wait for 4 minutes, add the content of one packet of

HI 93705B-0 Citric acid reagent and shake until it is
completely dissolved.

• Wait for 1 minute. This is the blank.