Hanna Instruments HI 83209-01 User Manual

Page 27

Advertising
background image

27

Chromium VI HR

CHROMIUM VI HIGH RANGE

SPECIFICATIONS

Range

0 to 1000 μg/L

Resolution

1 μg/L

Accuracy

±5 μg/L ±4% of reading at 25 °C

Typical EMC

±1 μg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 525 nm

Method

Adaptation of the

ASTM Manual of Water and Environmental Technology, D1687-92,

Diphenylcarbohydrazide method. The reaction between chromium VI and the reagent
causes a purple tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93723-0

Powder reagent

1 packet

REAGENT SETS

HI 93723-01 Reagents for 100 tests
HI 93723-03 Reagents for 300 tests
For other accessories see page 65.

MEASUREMENT PROCEDURE

• Select the

Chromium VI HR method using the procedure described in

the

Method Selection section (see page 12).

• Fill the cuvette up to the mark with 10 mL of unreacted sample and

replace the cap.

• Place the cuvette into the holder and close the lid.

• Press the Zero key. The meter will show “-0.0-” when the meter is zeroed and ready for measurement.

• Remove the cuvette and add the content of one packet of

HI 93723-0 reagent. Replace the cap and shake vigorously
for about 10 seconds.