Copper HR

• Remove the cuvette.

• Add the content of one packet of HI 93702-0 Bicinchoninate.

Replace the cap and shake gently for about 15 seconds.

• Reinsert the cuvette into the instrument.

• Press Timer and the display will show the countdown prior to the measurement or, alternatively, wait

for 45 seconds and press Read. When the timer ends the meter will perform the reading.

• The instrument displays the results in mg/L of copper.

INTERFERENCES

Interference may be caused by:
Silver
Cyanide
For samples overcoming buffering capacity of reagent (around pH 6.8), pH should be adjusted between
6 and 8.