Hanna Instruments HI 83200C User Manual

Page 29

Advertising
background image

56

57

Fluoride

FLUORIDE

SPECIFICATIONS

Range

0.00 to 2.00 mg/L

Resolution

0.01 mg/L

Accuracy

±0.03 mg/L ±3% of reading at 25 °C

Typical EMC

±0.01 mg/L

Deviation
Light Source

Tungsten lamp with narrow band interference filter @ 575 nm

Method

Adaptation of the

Standard Methods for the Examination of Water and Wastewater,

18

th

edition, SPADNS method. The reaction between fluoride and the liquid reagent

causes a red tint in the sample.

REQUIRED REAGENT

Code

Description

Quantity

HI 93729-0

SPADNS Reagent

4 mL

REAGENT SETS

HI 93729-01 Reagents for 100 tests
HI 93729-03 Reagents for 300 tests
For other accessories see page 128.

MEASUREMENT PROCEDURE

• Select the

Fluoride method using the procedure described in the

Method Selection section (see page 12).

• Add 2 mL of HI 93729-0 SPADNS Reagent to two cuvettes.

• Fill one of the cuvettes with distilled water up to the mark,

replace the cap and invert several times to mix.

• Fill the other cuvette with sample up to the mark, replace the

cap and invert several times to mix.

• Place the cuvette with the reacted distilled water (# 1) into the

holder and close the lid.

Cyanuric Acid

• Add the content of one packet of HI 93722-0 Cyanuric Acid

Reagent. Replace the cap and shake gently for about
10 seconds (dissolution is not complete).

• Reinsert the cuvette into the instrument.

• Press

Timer and the display will show the countdown prior to the measurement or, alternatively, wait

for 45 seconds and press

Read. When the timer ends the meter will perform the reading.

• The instrument displays concentration in

mg/L of cyanuric acid.

# 1

# 1

# 2

10 mL

10 mL

2 mL