106

107

Phosphorus

• Add the content of one packet of HI 93706B-0 Phosphorus

Reagent B (Amino Acid) to the cuvette. Replace the cap
and shake gently until completely dissolved.

• Reinsert the cuvette into the instrument.

• Press

Timer and the display will show the countdown prior to the measurement or, alternatively,

wait for 5 minutes and press

Read. When the timer ends the meter will perform the reading. The

instrument displays the results in

mg/L of phosphorus (P).

• Press ▲ or ▼ to access the second level functions.
• Press the

Chem Frm key to convert the result in mg/L of phosphate (PO

4

3

¯) and phosphorus pentoxide

(P

2

O

5

).

• Press ▲ or ▼ to return to the measurement screen.

INTERFERENCES

Interference may be caused by:
Sulfide
Chloride above 150000 mg/L
Calcium above 10000 mg/L as CaCO

3

Magnesium above 40000 mg/L as CaCO

3

Ferrous iron above 100 mg/L

Phosphorus

PHOSPHORUS

SPECIFICATIONS

Range

0.0 to 15.0 mg/L

Resolution

0.1 mg/L

Accuracy

±0.3 mg/L ±4% of reading at 25 °C

Typical EMC Dev. ±0.2 mg/L
Light Source

Tungsten lamp with narrow band interference filter @ 525 nm

Method

Adaptation of the

Standard Methods for the Examination of Water and Wastewater,

18

th

edition, Amino Acid method. The reaction between phosphate and reagents

causes a blue tint in the sample.

REQUIRED REAGENTS

Code

Description

Quantity

HI 93706

A-0

Molybdate

10 drops

HI 93706

B-0

Amino Acid Powder

1 packet

REAGENT SETS

HI 93706-01 Reagents for 100 tests
HI 93706-03 Reagents for 300 tests
For other accessories see page 128.

MEASUREMENT PROCEDURE

• Select the

Phosphorus method using the procedure described

in the

Method Selection section (see page 12).

• Fill the cuvette with 10 mL of unreacted sample (up to the

mark) and replace the cap.

• Place the cuvette into the holder and close the lid.

• Press the

Zero key. The display will show “-0.0-” when the meter is zeroed and ready for

measurement.

• Remove the cuvette.

• Add 10 drops of HI 937O6A-0 Molybdate reagent.

10 mL