CHAPTER TWO

Periodic Test

1. Every thirty (30) days the ODORATOR 2 should be operated with the power switch on

and while not hooked up to a gas supply.

2. Wait about 30 seconds and then position the nose approximately 3/4 inch (20 mm)

above the exhaust port and sniff the exhaust.

3. If no odor is detected, turn the ODORATOR 2

“OFF”. The ODORATOR 2 has not

absorbed the odorant and is usable.

4. If an odor is detected, turn the ODORATOR 2

“OFF” and send it to the nearest Heath

Repair Center (Regional Office) listed on page 16. The odor is an indication that the
ODORATOR 2 has absorbed the odorant and must have its internal tubing replaced
and gas flow components cleaned.


The forms on pages 13 and 14 are provided to document the Periodic Tests and are as a
courtesy only. Company procedures will define how records are kept and in what format.


Calibration

For regulatory compliance, instrument calibration and maintenance a factory calibration is
recommended every 12 but not more than 15 months from the date of the previous factory
calibration.

The ODORATOR 2 will automatically start alerting the user 30 days from the actual

annual calibration due date. Only Heath factory or factory trained personnel should perform
ODORATOR 2 calibration. Factory recognized labels and seals maintain warranty.

Refer to page 17 for Heath repair and calibration centers.