Instruction Manual - CMP/CMA series

Response time (95%)

Zero offsets
(a) Response to 200 W/m² net thermal radiation (ventilated)
(b) Response 5 K/hr change in ambient temperature

Non-stability
percentage change in responsivity per year

Non-linearity
percentage deviation from the responsivity at 500 W/m² due to the change of irradiance within the range 100 W/m² to 1000 W/m²

Directional response (for beam radiation)
the range of errors caused by assuming that the normal incidence responsivity is valid for all directions when measuring from any
direction a beam of radiation whose normal incidence irradiance is 1000 W/m²

Spectral selectivity
percentage deviation of the product of spectral absorptance and spectral transmittance from the corresponding mean within 0.35 µm
and 1.5 µm

Temperature response
percentage deviation due to change in ambient temperature within an interval of 50 K

Tilt response
percentage deviation from the responsivity at 0° tilt (horizontal) due to change in tilt from 0° to 90° at 1000 W/m² irradiance

1

2

3a

3b

3c

3d

3e

3f

Response time (95%)

Zero offsets
(a)
(b)

Non-stability

Non-linearity

Directional response
(for beam radiation)

Spectral selectivity

Temperature response

Tilt response

Second Class

< 60 s

± 30 W/m²
± 8 W/m²

± 3.0%

± 3%

± 30 W/m²

± 10%

8%

± 5%

First Class

< 30 s

± 15 W/m²
± 4 W/m²

± 1.5%

± 1%

± 20 W/m²

± 5%

4%

± 2%

Second Class

< 18 s

< 15 W/m²

(1)

< 5 W/m²

< 1%

< 1%

< 20 W/m²

up to 80° zenith angle

< 3%

< 5%

(-10°C to +40°C)
interval of 50 K

< 1%

Secondary Standard

< 5 s

< 7 W/m²

(1)

< 2 W/m²

< 0.5%

< 0.2%

< 10 W/m²

up to 80° zenith angle

< 3%

< 1%

(-10°C to +40°C)
interval of 50 K

< 0.2%

Secondary Standard

< 5 s

< 7 W/m²

(1)

< 2 W/m²

< 0.5%

< 0.2%

< 10 W/m²

up to 80° zenith angle

< 3%

< 1%

(-20°C to +50°C)
interval of 70 K

< 0.2%

Secondary Standard

< 5 s

< 3 W/m²

(1)

< 1 W/m²

< 0.5%

< 0.2%

< 5 W/m²

up to 80° zenith angle

< 2%

< 0.5%

(-20°C to +50°C)
interval of 70 K

< 0.2%

First Class

< 18 s

< 15 W/m²

(1)

< 4 W/m²

< 1%

< 1%

< 20 W/m²

up to 80° zenith angle

< 3%

< 4%

(-10°C to +40°C)
interval of 50 K

< 1%

Secondary Standard

< 15 s

± 7 W/m²
± 2 W/m²

± 0.8%

± 0.5%

± 10 W/m²

± 3%

2%

± 0.5%

1

2

3a

3b

3c

3d

3e

3f

Reference number

ISO Pyranometer Categories

Reference number and Specification

CMP10, 11/CMA 11

CMP 22

CMP 21

CMP 6/CMA 6

CMP 3

ISO 9060:1990 classification

Appendix B. Pyranometer classification to ISO 9060:1990(E)

(1)

without ventilation