Downflow velocities, Downflow test specifications – Labconco PURICARE Procedure Station Models 31260 User Manual

Page 85

Advertising
background image

Product Service 1-800-522-7658

85

Downflow Velocities

All Logic+ models, EXCEPT 30248xx80) are classified as having uniform downflow, and

the average downflow velocity for all models is 55+/-5 FPM, tested as per ANSI/NSF

Standard 49:2012.

All readings should be in a plane 4 inches above the bottom edge of the sash, and the

UV Lamp, IV bar and all other accessories must be removed before performing

downflow test. The test grid dimensions are outlined in Table 5-4. When taking readings

in the back two rows, the anemometer probe should be pointing toward the nearest side

wall, as shown in figure 5-16. When reading the front row, the probe should point

outward, toward the front of the cabinet, as shown in Figure 5-17. The thermal

anemometer’s time constant should be set at 10-15 seconds for greater data stability.

Table 5-4

Downflow Test Specifications

Cabinet Width (Feet)

3

3

4

4*

4

5

5

6

6

Sash Height (inches)

8, 9

10, 12

8, 9

8

10, 12

8, 9

10, 12

8, 9

10, 12

Nominal Avg. Downflow

(FPM)

55+/-5

55+/-5

55+/-5

75+/-5

55+/-5

55+/-5

55+/-5

55+/-5

55+/-5

Downflow test grid # of

points

21

21

24

24

24

30

30

36

36

Test Grid (# rows x #

columns)

3 x 7

3 x 7

3 x 8

3 x 8

3 x 8

3 x 10

3 x 10

3 x 12

3 x 12

Test Grid Data (in

inches)

Grid distance from back & sides

6.0

6.0

6.0

6.0

6.0

6.0

6.0

6.0

6.0

Distance between rows

5.75

5.55

5.75

5.75

5.55

5.75

5.55

5.75

5.55

Distance between columns

4.08

4.08

5.21

5.21

5.21

5.39

5.39

5.50

5.50

- Model #s 30248xx8x Only

Figure 5-16