Chapter 6 maintaining your enclosure, Routine maintenance schedule – Labconco Protector Work Stations 3930021 User Manual

Page 38

Advertising
background image

Product Service 1-800-522-7658

34










Chapter 6

Maintaining Your Enclosure

Monitoring airflow and changing the carbon filters (if equipped) is the primary
maintenance required.

Review this chapter on maintenance for the following:
1. Routine Maintenance.
2. Speed control adjustment and setting the inflow face velocity.
3. Operating and calibrating the airflow monitors.
4. Determination of when to replace carbon filters and how to replace.
5. Calculating carbon filter life.
6. Fluorescent light replacement.
7. Motorized impeller replacement.
8. Speed control replacement.

Routine Maintenance Schedule

Weekly

Wipe down the interior surfaces of the enclosure with a disinfectant or

cleaner, depending upon the usage of the unit and allow to dry.

Using a damp cloth, clean the exterior surfaces of the enclosure,

particularly the front and top to remove any accumulated dust.

Operate the exhaust system, noting the airflow velocity through the

enclosure using a source of visible smoke. Airflow monitors are
recommended for constant monitoring.