Laboratory Ventilation Standards

Federal Register 29 CFR Part 1910
Non-mandatory recommendations from “Prudent Practices.”

Fume hoods should have a continuous monitoring device

Face velocities should be between 60-100 linear feet per minute (lfpm)

Average 2.5 linear feet of hood space per person

Occupational Health and Safety
U.S. Department of Labor
200 Constitution Avenue N.W.
Washington, DC 20210
(202) 523-1452
www.osha.gov


Industrial Ventilation-ACGIH

Fume hood face velocities between 60-100 lfpm

Maximum of 125 lfpm for radioisotope hoods

Duct velocities of 1000-2000 fpm for vapors, gasses and smoke

Stack discharge height 1.3-2.0 x building height

Well designed fume hood containment loss, <0.10 ppm

Industrial Ventilation, A Manual of Recommended Practice.