Installing or Replacing the Training Pads Cartridge
To install the training pads cartridge, follow these steps:
1. If a cartridge is currently installed in the Trainer,

locate the latch at the top edge of the Trainer, and
slide it to the side.The pads cartridge will be
released. Remove the cartridge.

2. Remove the new training pads cartridge from its

resealable package.

3. Insert the M5073A or M5074A

Training Pads cartridge into the compartment on the front
of the HS1 Trainer (A).
Press the cartridge downfirmly until it clicks into place.
Be sure the green handle is pressed all the way down (B).
The Trainer will tell you what kind of training pads
-- Adult or Infant/Child --
have been inserted, then it allows you to select a training
scenario. If you do nothing, the Trainer will go into training
standby mode to be ready for use.

Selecting a Training Scenario
After battery installation
The default training scenario is number 1.
Each time batteries are installed in the Trainer, the training scenario
resets to the default.To use this scenario, install the training pads
cartridge, then simply pull the cartridge handle or press the On/Off
button.The Trainer will not announce the scenario number.

Selecting an alternative scenario
To select an alternative scenario, press the flashing blue i-button to scroll
through each of the eight scenarios in sequence.The Trainer will announce each
scenario by number as you press the button. When you reach the scenario
number you want, press and hold the On/Off button until the Trainer turns off,
or wait for 10 seconds and it will automatically go into training standby mode.
The Trainer will run the same selected scenario every time until the batteries
are replaced or until a new scenario is selected.

In training standby mode
If the i-button is pressed while the Trainer is in training standby mode, the devi-
ce will announce that it is training. If the i-button is pressed a second time, the
Trainer will identify the current training scenario number. A third press of the
flashing i-button will allow you to select another scenario.