Floor stand reactors – Parr Instrument Series 4530 User Manual

Page 7

Advertising
background image

Floor Stand Reactors

w w w . p a r r i n s t . c o m

7

Parr Heating Only:

The

Parr Heating output connector is to be used

only with Parr Instrument Company heater assem-
blies supplied with the appropriate heater power
cord.

Motor:

Secure the clamp on motor cord to the

controller with the provided screw next to

the motor socket for safety purposes.

The Motor output connector is to be used only with
Parr Instrument Company motor assemblies sup-
plied with the appropriate motor power cord.

2A. Vessels on movable cart.

The motor is contained in the overarm assembly
attached to the cart. The overarm assembly piv-
ots to the rear to allow the vessel to be placed in
the heater. The overarm is then aligned atop the
vessel. Pull up on the release knob to allow the
upper drive shaft to be connected to the mag-
netic stirrer.

2B. Vessels on stationary stand.

Bolt the stand to the fl oor using the

holes in the base plate.

Loosen the panel screw holding the vessel
retainer to the front midplate. Open the vessel
retainer to allow the placement of the vessel as-
sembly. If fi xed head style, position the head in
the slot located in the front midplate. If remov-
able head style, place vessel assembly such that
the split ring rests on midplate bracket. After
installing the vessel, close the vessel retainer
and retighten the panel screw. Pull up on the
release knob located at the top of the belt guard.

This knob is attached to the upper drive shaft
and the universal coupling. Lifting the knob will
allow the alignment of the universal coupling
and the drive adapter attached to the magnetic
drive. Turn the knob to align the slot in the drive
adapter with the universal coupling.

2.

3. Connect the power cord from the heater into the

heater socket on the rear panel of the controller.

4. Plug the motor power cord into the motor socket

on the rear of the controller.

Secure the clamp on the motor cord

with the provided screw next to the

motor socket for safety purposes.

5. Note the voltage requirement on the controller

identifi cation label on the back panel, and then
plug the power cord into an appropriate outlet.
Power for these reactors should be drawn from
a 3-slot, grounded outlet capable of carrying up
to the full current rating of the system.

6. Connect the thermocouple extension wire to

both the thermocouple and to the controller in
the “Primary Temp Input” position on the rear
panel. Insert the thermocouple into thermowell.

7. Connect leads from accessory packages such as

tachometer, pressure transducer and high temp
cut-off to the designated positions on the back
panel of the 4848 Controller.

8. Connect cooling water to internal cooling coil if

installed.

9. Connect cooling water to the magnetic stirrer.

See manual no. 234M.

10. Connect tubing to the rupture disc outlet and run

to a safely vented area. See manual no. 231M.

11. Using the switch on the Temperature Controller,

turn on the motor for a short run to check the
stirrer drive system.

Advertising