Identify the valves, Gas inlet valve, Gas release valve – Parr Instrument Series 4550 User Manual

Page 10: Liquid sampling valve, 4550 floor stand reactors

Advertising
background image

4550 Floor Stand Reactors

P a r r I n s t r u m e n t C o m p a n y

10

IDENTIFY THE VALVES

Gas Inlet Valve

The gas inlet valve is easily identified when the bomb
is open by noting that it is connected to a dip tube
that extends to a point near the bottom of the bomb
cylinder. This is an angle valve with an attached fitting
which provides a socket for attaching the A495HC
pressure hose furnished with the reactor.

Gas Release Valve

The gas release valve is connected to an opening
that will not have any fitting installed on the un-
derside of the head. Gas released from this valve
will be drawn from the top of the reactor. The gas
release port does include a 1/8

th

NPT thread on the

underside of the head where an additional dip tube
or catalyst addition device may be added. If one of
these additions is made to this port then an ad-
ditional valve must be added to the gage adapter to
be used as a gas release valve.

Liquid Sampling Valve

The liquid sampling valve is attached to the same
fitting as the gas inlet valve and connected to the
same dip tube. With this arrangement, incoming gas
is always introduced below the surface of the liquid
and the operator is provided with a means for clearing
the dip tube to be sure that any sample taken during
a run will be representative of the charge. This can be
done by opening the gas valve momentarily to force
any liquid in the dip tube back into the reactor before
withdrawing a sample from the sampling valve.

(COOLING COIL)

VALVE, LIQUID SAMPLING

(2) ELBOW,COOLING COIL

MALE CONNECTOR, DIP TUBE

GAGE ADAPTER

COMPRESSION RING

HEAD, REMOVABLE

(INLET/OUTLET)

GAS RELEASE VALVE

RUPTURE DISC ASSY

VALVE ADAPTER

DIP TUBE

PRESSURE GAGE

ADAPTER BUSHING

MAGNETIC DRIVE ASSY

ADAPTER BUSHING

VALVE ADAPTER

HEX PLUG (EXTRA PORT)

STIRRER SHAFT ASSY

THERMOWELL

SERPENTINE COOLING COIL

(2) IMPELLER ASSY

STIRRER BRACKET ASSY

THERMOCOUPLE

GAS INLET VALVE

ADAPTER, GAS CONNECTION

(2) MALE CONNECTOR

GLAND NUT, MAG DRIVE

Advertising