4550 Floor Stand Reactors

w w w . p a r r i n s t . c o m

11

OTHER VESSEL HEAD FITTINGS

Safety Rupture Disc

There is a safety rupture disc attached to the head that
is intended to rupture and release the pressure before
it reaches a dangerous level. A metal tag wired to
the safety head identifies the burst pressure at room
temperature for that particular disc. A similar tag is
furnished with each replacement disc. This tag must
remain with the apparatus at all times so that both pres-
ent and future operators will be aware of the disc rating.
Users should read the discussion of rupture discs given
in the Instruction Sheet No. 231M for a complete de-
scription of the characteristics of rupture discs and the
precautions to be observed when operating pressure
equipment protected by this type of safety device.

A typical pre-bulged disc can be used to 90% of the
rating on the tag. For additional protection, the user
should install an adequate and safe venting system for
removing any toxic, flammable or volatile material which
would be released if the rupture disc should burst. A
connector for attaching 3/8” OD tubing to the discharge
port of the rupture disc is provided for this purpose.

Type J Thermocouple

A Type J thermocouple in a 1/8” dia. stainless steel
sheath is furnished with the reactor. Insert this ther-
mocouple into the head thermowell and connect
it to the thermocouple socket on the rear panel of
the temperature controller using the extension wire
furnished with the reactor.

Pressure Gage

The pressure gage furnished with this reactor has a T316
Stainless Steel Bourdon tube. Gages are furnished in
a variety of ranges to meet individual needs. Typically,

the gage and the rupture disc are furnished as matched
ranges. For applications where a gage is selected with
a range under 1000 psi, a relief valve is added and set
to protect the gage. A 1000 psi rupture disc is installed
as the fail-safe vessel protection.

For highly corrosive applications where the vapor
phase might corrode the stainless Bourdon tube, Parr
offers isolator assemblies in a variety of materials.
These isolators with their internal piston isolate the
vapors from the gage.

The gage adapter includes a ¼” NPT side port with a
plug installed. This position may be used for a variety
of fittings such as a needle valve, pressure transducer
or relief valve.

Gage and Valve Adapters

The pressure gage and the combined gas inlet and
sampling valves are attached to the head with an
adapter which allows these fittings to be drawn up
tightly when facing in any direction. To attach these
fittings to the head, screw the gage or valves firmly
into the adapter, then run the 209HC4 bushing onto the
threaded stem as far as it will go. Screw this assembly
into the head until the nose of the adapter is seated;
then back it off until the valve or gage is facing in the
desired direction. Now hold the fitting firmly in place
and close the joint by tightening the 209HC4 bushing.
This connection can be made and broken repeatedly
without destroying the sealing faces. A light coating of
thread lubricant, such as Parr No. 424HC2 High Tem-
perature Anti-Seize Lube, applied to the threads and to
the nose of the adapter will help to obtain a tight joint.

Note:

Do not use PTFE tape on the straight thread

connections of the coned adapters and mating

bushings. PTFE tape should only be used on the (NPT)

threads such as the needle valves or gage connection.

208HC15

PRESSURE GAGE

56HC

209HC4

GAS RELEASE VALVE

658HC

A130VB

64

DO NOT SCALE DRAWING

REVISIONS

COMPANY.

APPROVED

BY

BY

SCALE

DWG NO.

DATE
DRAWN

OR BETTER

DESCRIPTION

DRAWN IN 3rd ANGLE PROJECTION

DATE

1:2

EWL

211 53rd STREET

07-22-13

PARR INSTRUMENT CO.

4550

MOLINE, ILLINOIS 61265

SIZE

B

FOR

395M MANUAL, 4550 FLOOR REACTOR

TOLERANCES

IN INCHES

UNLESS OTHERWISE SPECIFIED

1/X .... ±1/64

.00 .... ±.010

.000 .... ±.003

ANGULAR ... ±1/2°

UNMARKED RADII .03"

MACHINED SURFACES:

PROPRIETARY

NEITHER THE DRAWING

NOR INFORMATION

CONTAINED HEREIN

MAY BE COPIED,

REPRODUCED, OR

OTHERWISE USED

WITHOUT WRITTEN

PERMISSION FROM

PARR INSTRUMENT

395M_S10

A131VB

420HC2

209HC4

INLET SAMPLING VALVE

A129VB

208HC6