Mtm installation, Module – Parr Instrument MTM User Manual

Page 6

Advertising
background image

MTM Installation

P a r r I n s t r u m e n t C o m p a n y

6

Module

Keys command:
1. Press “SET” to select
2. Press return key move to next operation mode
3. Up/Down arrow keys to adjust value or select

type


Main Screen: SV = 100

Press return key and release

Operation

Mode

Select

type/value Comment

r-S

Run

Run/Stop

SP

1

Decimal point position

LoC

OFF

Lock mode (lock all
keys or only up/down
arrow able to use)

Press and hold down “SET” for 5-sec

Operation

Mode

Select

type/value Comment

InPt

v10

Input type (v5 = 0-5V,
v10 = 0-10V, nA0 =
0-20mA, nA4 = 4-20mA,
nV =0-50mV)

tP-H

To be de-
termined*

Upper-limit range

tP-L

0

Lower-limit temperature
range

CTRL

ON/OFF

Control mode (ON/OFF,
MANUAL, PID and PID
PROG)

S-HC

Heat

Heat/Cool control

ALA1

0

Alarm mode

SALA

OFF

System alarm

CoSH

ON

Communication write
function that able to use
set point from software

C-Sl

ASCII

Format type

C-no

See note** Communication address

bPS

9600

Communication baud
rate

Len

7

Data length setting

Operation

Mode

Select

type/value Comment

PrtY

Even

Parity bit setting

StoP

1

Stop bit setting


* For 90VDC motor, 115VAC, tP-H = 142.9

For 180VDC motor, 230VAC, tP-H = 120.4

For 90VDC motor, 230VAC, tP-H = 153.2

** First available communication port (3 or 4)

Press “SET” and release

Operation

Mode

Select

type/value

Comment

HtS

0

Hysteresis

TPoF

0

Inaccuracy adjustment

CrHI

0

Analog output low

CrLo

0

Analog output high

Advertising