REF-METER-OCTA User’s Guide

English

10

Tank Capacity Mode Training

1. While

holding

down, press

. The SET TANK indicator will turn on, and OFF will

show on the display. This indicates that the Tank Capacity Mode is currently off.

2. Turn the Tank Capacity Mode on by pressing . The display will show ON, and the

Alternate Numeric Display and Percent Indicator will turn on.

3. Press

to proceed to the first step of training. 0% will flash in the Alternate Numeric

Display indicating that the weight of an empty container must be registered.

4. Apply an empty container (or a load representing the weight of an empty container) on

the platform. Wait 5 seconds for the load to stabilize, then press

. 100% will flash in

the Alternate Numeric Display indicating that the weight of a full container must be
registered.

5. Apply a full container (or a load representing the weight of a full container) on the

platform. Wait 5 seconds for the load to stabilize, then press

. This completes the

training process, and the SET TANK indicator will turn off.

6. From this point forward, the Alternate Numeric Display will show the percent of material

remaining in the container from 0% to 100%.

NOTE: The 100% weight must be larger than the 0% weight by at least 200g. If not, “Err
will show on the Alternate Numeric Display. To clear the error, retrain as above using a
100% weight that is at least 200g larger than the 0% weight. During training, the ZERO key
may be used to assist in setting the 0% and 100% values. This will have no effect on the
training, as the Tank Capacity Mode relies on absolute, not relative, weights.