4. CONNECTIONS

This module operates only when connected to a PC USB bus, using the

included type-A to mini-B cable. The following figure shows all possible input
connections.

Warning: Signal wires connected to all inputs should be kept separated from power

wires, and when possible installed inside of grounded electrical ducts.



3.1 Pt100 Wiring

Connections of channel 1 are on terminals 6, 7 and 8. For channel 2

on terminals 2, 3 and 4. Connection with 3 wires from the sensor element to the

module is necessary to cancel the cable resistance error in the measurement.

The 3 wires must have the same length and gauge. To connect a Pt100 with 2

wires, connect terminals 7 and 8 (channel 1) or 3 and 4 (channel 2).


3.2 Thermocouple Wiring

Connections of channel 1 are on terminals 6 and 7. For channel 2 on

terminals 2 and 3. Observe indicated polarity. Cables for thermocouple

connection must have the same thermo-electrical characteristics of the

thermocouple element (compensation or extension cable). Observe that both the

thermocouple and the compensation cable must be connected with correct

polarity. If compensation cable is not used or is not connected with the correct

polarity, large measurement errors will arise.