American Weigh Scales (AWS) CG-500 User Manual

Page 4

Advertising
background image

6. Remove the sample quantity from the platform.

7. You now can place any amount of your item(s) on

the scale and the LCD will display a count.

8. You can switch between weighing mode and

counting mode by pressing [

MODE ] or [ PCS ] key

respectively.

IV. CALIBRATION

Calibration is NOT required before first use. Only

calibrate if the scale is not weighing properly.

The CG-500 calibrates with a 500g weight*

The CG-50 calibrates with a 50g weight*

*Weights not included.

1. Turn on the scale by pressing the [

ON/OFF ] key.

2. Wait until the LCD displays “0.0” or "0.00", then

press and HOLD the [

MODE ] key for 3 seconds. The

LCD will display “CAL”. You may now release the

[

MODE ] key.

3. Press the [

MODE ] key again. The LCD will display

“CAL” followed by the required calibration weight (see

above).

4. Place the required calibration weight on the center

of the platform. The display will show “PASS” and

return to normal weighing mode.

5. Put away the calibration weight and press

[

ON/OFF ] to turn off the scale and the calibration is

finished!

Advertising
This manual is related to the following products: