Hot-air sterilization at 180°C

Grenzert

Manual abort

of sterilization

3 sec.

Normal Display

Start the sterilization cycle

After the end of sterilization

Running sterilization

Turn off the unit,
or press “180°C”,
or open the outer door.
The unit is ready to operate.

Ineffective sterilization when aborting prematurely!

3 sec.

Do NOT open the doors until the
interior temperature has
dropped to 37°C. After a manual
abortion, the C 150 reverts to its
standard operational status. As
long as the interior temperature
remains above the temperature,
which was set to trigger the
temperature alarm, the C 150
will trigger alarm signals. You
can switch them off by pressing
“EXIT”. If necessary, repeat the
sterilization.

10 to 12 hours

Alternating display

Alternating display

showing the actual temperature

Copyright

BINDER GmbH 10/2012

Art. No. 7001-0235

©

Hot-air sterilization of the inner chamber

The hot-air sterilisation takes10 to
12 hours in 3 steps:
- Rapid heating-up
- Maintaining the inner

temperature for 4 hours,

on

- Cooling-down phase up to 37°C

which