Calibration reference conditions:

Free field perpendicular incidence,

Ambient temperature: 20

o

C,

Relative humidity: 65%,

Measurement Range: Hi (high),

Excitation SPL: 114 dB,

Frequency: 1 kHz.

Effect of humidity on accuracy:

Less than ±0.5dB over the range 30 to 90%

RH (provided there is no condensation),

relative to the value at 65% RH and 40

o

C.

Storage range: 20 to 95% RH

non-condensing.

Effect of temperature on accuracy:

Less than ±0.5dB over range -10 to +40

o

C.

For temperatures between +40 and +50

o

C

it is recommended that the instrument be

recalibrated at the operating temperature

with a CEL-282 or CEL-284/2.

However above 40

o

C, accuracy within the

limits of IEC 651 can be obtained using the

following correction, where t is

o

C:

Corrected reading = display reading + 0.05(t-40).

Operating temperature range: -10 to +50

o

C.

Storage range: -20 to + 60

o

C.

Effect of magnetic fields on accuracy:

Negligible at all levels.

Lower than a reading of 30 dB(A) when

tested according to IEC 651 clause 8.4.

Effect of vibration on accuracy:

Less than 3 dB when tested according to

IEC 651 clause 8.3.

Directional response:

Within the limits of IEC 651.

Effect of operator:

Negligible when instrument held at arm’s

length with microphone pointing away from

operator.

Frequency weightings:

A-weighting according to ANSI S1.4 Type

2A, and IEC 651 Type 2.

C-weighting according to ANSI S1.4 Type

2A, and IEC-651 Type 2.

Time weightings:

F (fast):

125 milliseconds,

S (slow):

1 second,

I

(impulse):

35 millisecond rise time,

1.5 second decay time, with

quick reset.

MAX:

Holds noise readings of

F, S, &

I

, with decay <1 dB /

5 minutes at temperatures

<40

o

C.

Digital display:

3

1

/

2

digits,

7 segments,

12.5 mm high characters,

0.1 dB resolution.

Display updated 3 times per second.

Functions indicated:

Sound level dB(A),

Low battery level. (Indicated when battery

falls below approximately 3.2 V).

Sound level dB(C),

Battery voltage (x10),

Overload,

Maximum SPL (A- or C-weighted).

Overload detection: (CEL-254 Only)

Pre-weighting overload detector, with 2.5 s

dwell time.

Start up stabilisation time:

20 seconds.

Outputs:

Two outputs from standard 3 pole 3.5 mm

coaxial socket with a.c on pin, d.c on

intermediate, and ground on sleeve.

DC: Log. signal with nominal 25 mV/dB

(offset by approx. 4 to 5 V) and not

exceeding 6 V,

125 ms rise & decay times for Fast,

1 s rise time & decay time for Slow.

35 ms rise time and 1.5 s decay time

for Impulse.

Recommended impedance of load

>27 kΩ.

AC: Conditioned signal, 7.25 V RMS for

FSD,

Recommended impedance of load

>39 kΩ.

Batteries:

4 x IEC AAA Manganese Alkaline type.

Battery life is typically better than 20 hours

with continuous use.

Dimensions:

258 x 70 x 21 mm, (10 x 2.75 x 0.85 inches).

Weight:

200 g, (0.45 lb) excluding batteries.

Compliance

The CEL-231 and CEL-254 Sound Level

Meters comply with the EMC Directive

89/336/EEC of the European Union. They have

been tested according to the standard delivery

schedule and comply with the following

standards.

EN 50081-1: 1992 & EN 50081-2: 1993

Generic emission standards for residential,

commercial, light industry and industrial

environments.

EN 50082-1: 1992 & EN 50082-2: 1995

Generic immunity standards (for both RF fields

and electrostatic discharge) for residential,

commercial, light industry and industrial

environments.

Manufacturers Servicing & Warranty Arrangements

In order to ensure its rigid conformity with

the requirements of the specification, this

instrument is thoroughly inspected and

calibrated prior to dispatch from the factory. All

technical information for an individual

instrument is filed under the instrument serial

number. Therefore, the serial number should be

quoted in any correspondance concerning the

instrument.

The manufacturers undertake to rectify any

defect in the instrument that is directly

attributable to faulty design or assembly, and

which becomes apparent during the warranty

period. In order to take advantage of this

warranty, the instrument must be returned,

carriage paid, to the manufacturer’s factory or

accredited agent, where necessary repairs will

be carried out.

Normally, the warranty period runs for 12

months from the date of receipt of goods, with

exceptions on certain specialised components

supplied by other manufacturers which are

warranted for shorter periods. Some of the

specialised components used in this instrument

may be subject to longer guarantees by their

actual manufacturers, and in all such cases, the

benefit of these undertakings will be passed on

to the user. However, Casella CEL Ltd’s liability

is limited to items of their own manufacture,

and they do not accept liability for any loss

resulting from the operation or interpretation of

the results from this equipment.

A comprehensive Instrument Calibration

Maintenance Agreement (ICMA) scheme is

available to extend the initial warranty period of

this instrument. At the end of the first warranty

period, it is recommended that the equipment

be returned to the Service and Re-calibration

Department at Bedford, where it will be

inspected and entered into the ICMA scheme

as required. The warranty will then be extended

for the period stated on the individual schedule.

Please contact your local Casella CEL agent for

full details of this service.

In the event of a malfunction developing

during the warranty period, the instrument

should be carefully packed and returned either

to Casella CEL’s local agent, or in the case of

domestic sales, to the Service Department at

Bedford.

Please include the following information:

Instrument Type(s) and Serial Number(s),

Customer name and address,

Contact name and phone number,

Reason for returning the equipment with a

detailed description of the fault.

The necessary adjustments or repairs will

be carried out, and the instrument returned as

soon as possible. After the warranty has expired

(except on approved accounts) service work is

undertaken against quotations, and all packing

and transit costs are charged extra.

CEL and Dawe instrumentation is designed, manufactured, and serviced by:

CASELLA CEL

Regent House, Wolseley Road, Kempston, Bedford, MK42 7JY, U.K

Telephone: +44 (0) 1234 844 100 Fax: +44 (0) 1234 841 490 e-mail; [email protected]

CASELLA CEL

Tech Park, 17 Old Nashua Road, Box 15, Amherst, NH 03031, U.S.A.

Office Phone: +1 603 672 0031 Toll Free: 1 800 366 2966 Fax: +1 603 672 8053

e-mail: [email protected]

Think environment think Casella

www.casella.co.uk

DIGITAL IMPULSE SOUND LEVEL METER

dB

CAL

HI = 65-135 dB(A)

LO= 30-100 dB(A)

RANGE

MODE

RESPONSE

A HI
ALO

C LO

SPL

MAX
BAT

C HI

OFF

F
S

IR

I

01013

1/4" Electret Microphone

Range Switch
(ONLY A HI & A LO ON CEL-231)

Response / Power Switch
(ONLY F & S ON CEL-231)

Retaining Clip (2)

Calibration Control

Coloured Cover

Standard 3.5 mm 3- Pole
Coaxial Output Socket

(NOT FOR POWER)

Overload Indication
(NOT CEL-231)

Display

Mode Switch
(NOT CEL-231)

CEL-254

CEL-254 DIGITAL IMPULSE

SOUND LEVEL METER

CEL-231 DIGITAL

SOUND SURVEY METER

Instructions

060064

Issue 8

Jan 02

CEL and Dawe Equivalent Identities

Instrument

Class 1L

Calibrator

Class 2L

Calibrator

Digital Sound

Survey Meter

Digital

Impulse SLM

Digital

Integrating SLM

Personal Sound

Exposure Meter

CEL Identity

CEL-284/2

CEL-282

CEL-231

CEL-254

CEL-269 (Obs)

CEL-272 (Obs)

Dawe Identity D-1418D

D-1411E

D-1405E

D-1422C

D-1421D (Obs)

D-1423D (Obs)

Advertising
This manual is related to the following products: