Symptom:- Pump runs but cannot achieve desired flow rate.

Possible cause:-

Service Hint.


Leak on pump assembly.

Locate and correct.

Dirt under valves or valves damaged.

Replace.

Inlet filter clogged.

Remove and clean.

Flow potentiometer inoperative

Replace circuit

board.

Symptom:- Cannot obtain full shift
working even when pressure drop does
not exceed 40cm H20 at 2 litres/min.

Possible cause:-Incomplete charge.

Recharge.

Battery below capacity.

Fully charge and

check voltage, should

be 4.8V-5.4V. If

below check output

from charger:-

should be 110-125mA.

If output correct

replace battery.
High current consumption due to

Locate and correct.

mechanical defect in pump.