5.

MAINTENANCE & SPARES

5.1

Routine Care

No routine maintenance is required.

Do not oil spindle pivots as this attracts dust which may make the

movement slow and can cause inaccuracy. If the instrument is used in a

dusty environment use a fine brush to remove any dust from the hair and

linkage pivots.

Do not wash the hair with water, except at long intervals or for special

reasons, as wetting the hair will upset the calibration temporarily.

5.2

Quartz Clock

The quartz clock should not require any attention, apart from battery

replacement. Quartz movements contain no parts that can be serviced by

the user. If the clock fails, a replacement must be obtained from Casella at

Bedford.

5.3

Cleaning

Cleaning may be carried out by using a soft damp cloth. Solvents or other

cleaning products should not be used.

5.4

Service

The original setting of the temperature element will remain correct for a long

period and will only need attention in the case of damage. The humidity

element does not have such a long life and will need replacement. The

interval depends on exposure and general care during use.

Indication of the need for replacement is when checks on accuracy show

large errors, when re-adjustment is needed very frequently, or erratic

hysteresis errors are noted.

When a temperature check and a humidity check are necessary, the

temperature check must be carried out first.

Casella's in house service department offers a comprehensive range of repair

and calibration services, designed to effect a fast and efficient back-up for all

our products. The Service Department is operated under the scope of our BSI

registration for products manufactured by us. We will however, undertake the

repair of other manufacturers equipment.

Maintenance

Page 9 of 12

THERMOHYGROGRAPH deluxe
User Manual