Microphone type

Coupler required

Correction to nominal level dB

CEL-186/2P CEL-2835

-1.0

CEL-186/2PR CEL-2835

-1.0

CEL-192F CEL-2835

-1.0

CEL-190F Not

required

-1.0

Genrad 1961

Not required

0

Genrad 1962

CEL-2834

0

Knowles BL1803

CEL-2834

0

B&K 4133

CEL-2835

-1.0

Figure 1

Microphone equivalent volume corrections to nominal output sound level

CEL-177 Acoustic Calibrator

level variation with altitude

-6.00

-5.00

-4.00

-3.00

-2.00

-1.00

0.00

1.00

2.00

-5 -4 -3 -2 -1 0 1 2 3 4 5 6 7 8 9 1011121314151617181920

Altitude (ft x 1000)

Correction to output

level (dB)

Figure 2

Calibration corrections as a function of altitude realtive to sea level

CEL Dosimeter

model number

Dosimeter calibration count at

nominal output 114 dB

(includes coupler where used)

CEL-139D

58.2

CEL-172H & J

58.2

CEL-172K

184.8

CEL-172L

116.4

CEL-179H & J

46.2

CEL-179K

146.7

CEL-179L

92.4

CEL-180G

46.2

CEL-180J

146.7

CEL-180L

5.5

CEL-181G

58.2

CEL-181J

184.8

CEL-181L

6.3

Figure 3

Dosimeter calibration settings using CEL-177 Acoustic Calibrator