Technical Description Concerning Electromagnetic Emission

Table 1: Declaration - electromagnetic emissions

Guidance and manufacturer’s declaration – electromagnetic immunity

The model TL30 is intended for use in the electromagnetic environment specified below.

The customer or the user should assure that It is used in such an environment.

Immunity test

IEC 60601

test level

Compliance

level

Electromagnetic environment –

guidance

Electrostatic

discharge (ESD)

lEC 61000-4-2

±6 kV contact

±8 kV air

±6 kV contact

±8 kV air

Floors should be wood, concrete

or ceramic tile. If floors are covered

with synthetic material, the relative

humidity should be at least 30 %.

Electrical fast

transient/burst

IEC 61000-4-4

±2 kV for power

supply lines

±1 kV for input/output

lines

±2 kV for power

supply lines

Mains power quality should be that

of a typical commercial or hospital

environment.

Surge

lEC 61000-4-5

±1 kV line to line

±2 kV line to earth

±2 kV line to earth

Mains power quality should be that

of a typical commercial or hospital

environment.

Voltage dips,

short interrup-

tions and vol-

tage variations

on power supp-

ly input lines

IEC 61000-4-11

<5 % U

T

(>95% dip in U

T

)

for 0.5 cycle

40 % U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

<5% U

T

(>95 % dip in U

T

)

for 5 sec

<5 % U

T

(>95% dip in U

T

)

for 0.5 cycle

40 % U

T

(60% dip in U

T

)

for 5 cycles

70% U

T

(30% dip in U

T

)

for 25 cycles

<5% U

T

(>95 % dip in U

T

)

for 5 sec

Mains power quality should be that

of a typical commercial or hospital

environment. If the user of the

model TL30 require continued ope-

ration during power mains interrup-

tions, it is recommended that the

model 168 DAYLUX Therapy Light

be powered from an uninterruptible

power supply or a battery.

Power frequen-

cy (50/60 Hz)

magnetic field

lEC 61000-4-8

3 A/m

Not applicable

Not applicable

NOTE U

T

is the a.c. mains voltage prior to application of the test level.