Table 3

Guidance and manufacturer’s declaration – electromagnetic immunity
The FT78 Ear thermometer is intended for use in the electromagnetic environment specified below. The customer or the user of the FT78

Ear thermometer should assure that it is used in such an environment.
IMMuNITY

test

IEC 60601

test level

Compliance

level

Electromagnetic environment – guidance

Conducted RF

IEC 61000-4-6

Radiated RF

IEC 61000-4-3

3Vrms

150 kHz to

80 MHz

3 V/m

80 MHz to

2.5 GHz

3Vrms

3 V/m

Portable and mobile RF communications equipment should be used no closer to any part of

the FT78 Ear thermometer, including cables, than the recommended separation distance cal-

culated from the equation applicable to the frequency of the transmitter.
Recommended separation distance:
d = 1.2

d = 1.2

80 MHz to 800 MHz

d = 2.3

800 MHz to 2.5 GHz

Where P is the maximum output power rating of the transmitter in watts (W) according to the

transmitter manufacturer and d is the recommended separation distance in meters (m).
Field strengths from fixed RF transmitters, as determined by an electromagnetic site survey,

a

should be less than the compliance level in each frequency range.

b

Interference may occur in the vicinity of equipment marked with the following symbol:

NOTE 1 At 80 MHz and 800 MHz, the higher frequency range applies.

NOTE 2 These guidelines may not apply in all situations. Electromagnetic propagation is affected by absorption and reflection from struc-

tures, objects and people.

Field strengths from fixed transmitters, such as base stations from radio (cellular/cordless) telephones and land mobile radios, amateur

radio, AM and FM radio broadcast and TV broadcast can not be predicted theoretic call with accuracy. To assess the electromagnetic envi-

ronment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in

which the FT78 Ear thermometer is used exceeds the applicable RF compliance level above, the FT78 Ear thermometer should be observed

to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as reorienting or relocating

the FT78 Ear thermometer.

Over the frequency range 150kHz to 80MHz, field strengths should be less than 3 V/m.