Table 2

Guidance and manufacturer’s declaration – electromagnetic immunity

The FT70 Multi-function Thermometer is intended for use in the electromagnetic environment specifi ed below. The customer or the user
of the FT70 Multi-function Thermometer should assure that it is used in such an environment.

IMMUNITY test

IEC 60601 test level

Compliance level Electromagnetic environment – guidance

Electrostatic discharge (ESD)
IEC 61000-4-2

± 6 kV contact
± 8 kV air

± 6 kV contact
± 8 kV air

Floor should be wood, concrete or ceramic tile. If fl oors are
covered with shythetic material, the relative humidity should
be at least 30%.

Electrical fast transient/burst
IEC 61000-4-4

±2 kV for power
supply lines
±1 kV air for input/out-
put lines

Not applicable

Mains power quality should be that of a typical commercial
or hospital environment.

Surge
IEC 61000-4-5

±1kV differential mode
±2kV common mode

Not applicable

Mains power quality should be that of a typical commercial
or hospital environment.

Voltage dips, short interrup-
tions and voltage variations
on power supply input lines
IEC 61000-4-11

<5% U

T

(>95% dip in

U

T

for 0.5 cycle

40% U

T

(60% dip in

U

T

for 5 cycles

70% U

T

(30% dip in

U

T

for 25 cycles

5% U

T

(>95% dip in

U

T

for 5 sec

Not applicable

Mains power quality should be that of a typical commer-
cial or hospital environment. If the use of the FT70 Multi-
function Thermometer requires continued operation during
power mains interruptions, it is recommended that the FT70
Multi-function Thermometer be powered from an uninter-
ruptible power supply or a battery.

Power frequency (50/60 Hz)
magnetic fi eld IEC 61000-4-8

3 A/m

3 A/m

Power frequency magnetic fi elds should be at levels cha-
racteristic of a typical location in a typical commercial or
hospital environment.

NOTE: U

T

is the a.c. mains voltage prior to application of the test level.