13.0 Cleaning the conductivity/temperature probe

Cleaning the Bluelab Conductivity/Temperature Probe

Cleaning the conductivity/temperature probe periodically ensures accurate readings.

The conductivity/temperature probe is cleaned using the Bluelab Conductivity Probe Cleaner, or “Jif”
a trade name for a liquid scourer cream used in home bathrooms and kitchens. Similar products are
called “Liquid Vim”, “Soft Scrub”, “Cif cream”, or “Viss”. Never use scented varieties as they contain
oils that contaminate the conductivity/temperature probe.
Follow the steps below to clean the conductivity/temperature probe.

Remove shroud.
Warm the shroud in your hand for a few

seconds to help with removal. Hold the

body and pull the shroud off.

Clean the conductivity probe face.
Place one or two drops of Bluelab

Conductivity Probe Cleaner onto the probe

face and rub with the Bluelab Chamois or

your fi nger fi rmly and vigorously.

Rinse the conductivity probe face.
Rinse off all traces of cleaner under

running tap water while scrubbing the

probe face with the other side of the

Bluelab Chamois or the same fi nger.

Check that the water forms a smooth

fi lm on the probe face. Ensure you have

a clean, smooth fi lm without any beads

of water.
If you have beads of water, repeat steps 2

and 3.

Refi t the shroud and test in 2.77 EC

Conductivity Standard Solution to

ensure adequate cleaning.
Place the probe tip into the solution, wait

for the reading to stabilize to a constant

value. This can take a few minutes while

the probe adjusts to the temperature of

the solution.
Repeat the cleaning process if the reading

given is not within 0.1 EC, 1 CF, 50 ppm or

70 ppm of the values in the table below.

Testing the Bluelab Conductivity/Temperature Probe

The conductivity/temperature probe is tested in either
2.77EC/27.7CF/1385 ppm or 1940 ppm solution
depending on the unit of conductivity chosen.

Use the standard solutions in the table to the
right. Bluelab solutions are recommended.
NOTE: The shroud MUST be left on the probe when taking readings.

1

2

3

4

5

Unit

chosen

EC CF ppm

500

ppm

700

Solution
required

2.77 27.7 1385 1940

uneven beads

of water

2.77 EC Conductivity

Standard Solution

clean, smooth

fi lm of water

1

3

3

5

2