0 ph calibration cont, English – Bluelab Guardian Monitor Connect Manual User Manual

Page 9

Advertising
background image

English

9

10.0 pH calibration cont.

To calibrate the pH

Clean pH probe tip.
See section 14.0 (the pH probe does not

require cleaning before the fi rst use).

pH 7.0 calibration
a) Prepare a small amount of pH 7.0

calibration solution in a container.

b) Ensure the Guardian Monitor Connect

is plugged in and in operation mode.

c) Rinse the pH probe tip in fresh water,

shake off excess water and place in the

pH 7.0 calibration solution. Wait at least

one minute for reading to stabilize.

d) Press and hold the pH cal button for a few

seconds until ‘PH’ and ‘CAL’ appear in the

windows, then release the pH cal button.

The ‘7’ LED will glow green. All three

windows will return to monitoring mode

with all current values being displayed.

e) If ‘Err’ is displayed see the

troubleshooting guide on section 16.0.

pH 4.0/10.0 calibration
a) Prepare a small amount of pH 4.0 or

pH 10.0 calibration solution in a container.

b) Rinse the pH probe tip in fresh water,

shake off excess water and place in

either pH 4.0 or pH 10.0 calibration

solution. Wait at least one minute for

reading to stabilize.

c) Press and hold the pH cal button for a few

seconds until ‘PH’ and ‘CAL’ appear in the

windows, then release the pH cal button.

The 4/10 LED will glow green. All three

windows will return to monitoring mode

with all current values being displayed.

d) If ‘Err’ is displayed see the

troubleshooting guide on section 16.0.

e) The Guardian Monitor Connect is now

calibrated and ready for use.

1

2

3

Successful pH 7 and pH 4 calibration

pH 4.0 or

pH 10.0

calibration

solution

pH 7.0

calibration

solution

clean the pH probe

rinse in tap water

Advertising