6 in-line filter, 7 fluid connections, 8 electrical connections – Bronkhorst CEM W-300B User Manual

Page 9

Advertising
background image

BRONKHORST HIGH-TECH B.V.

2.6 In-line filter

Although fluids and piping should be absolutely free from dirt, it is recommended to install inline filters

upstream of the flowmeter(s).

2.7 Fluid connections

The liquid meter should be installed according to its manual. The outlet of this meter should be connected to

the liquid input port of the CEM valve. The controller section of the liquid flowmeter controls the CEM valve.

Electrical connections should be made according to the customer system description.

The outlet of the gas flow controller must be connected to the gas input part of the CEM valve.

The vapour outlet of the CEM should be directly coupled to the process, this tube should be heated at a

temperature higher than the heater in order to assure that the vapour will not condense in this line. Avoid

“cold spots” in the outlet.

* Note: Always check your system for leaks, before applying fluid pressure. Especially if toxic, explosive or

other dangerous fluids are used.

2.8 Electrical connections

2.8.1 General

Bronkhorst HIGH-TECH B.V. recommends using their standard Readout and Control module and standard

cables, which have been tested together with the system. These cables are always supplied with the CEM

system.

When not using the standard Readout and Control unit please read chapter 4.

2.8.2 Pin configuration

The heat exchanger is equipped with two connectors:

connector (1) 6-pin male

Pinnumber

Description

A

B

C

D

E

F

Not connected

Not connected

PT100 (T-sensor)

PT100 (T-sensor)

Safety switch

Safety switch

connector (2) 5-pin male

Pinnumber

Description

110…120Vac supply

Description

220…240Vac supply

A

B

C

D

E

PE

Heater (N)

Heater (L)

Heater (L)

Heater (N)

PE

Not connected

Heater (N)

Heater (L)

Not connected

9.17.082

page 9