Menu 3. scale calibration settings, Calibration – Digi-Star SLC 2400 User Manual

Page 18

Advertising
background image

Menus and Calibration

18

SLC 2400 Operator’s Manual

D3924

SETTING

[display]

D.A.N

NO.

OPTIONS [displayed]

BOLD=DEFAULT

DESCRIPTION

REMOTE

(REMOTE)

218

ON

/OFF

If ON scale communicates

with cab control display

ZERO OUTPUT

(zerout)

219

Perform Zero/Balance for

SCOREM #11 weight output

and analog output (4-20mA)

SCALE NUMBER

(scl no)

231

Select scale number for cab

control communication

REMOTE

DISPLAY

(rmdisp)

234

EZ3MUX

EZ2

Select type of remote display

ANALOG LOW

WEIGHT

(LOW WT)

241

Enter analog weight value to

equal 4mA or 0 volts

ANALOG HIGH

WEIGHT

(HIGHWT)

242

Enter analog weight value to

equal 20mA or 5 volts

ANALOG SELECT

(ANAOUT)

243

Select 0-5V,4-20ma or 0-20ma

output

MENU 3. SCALE CALIBRATION SETTINGS

DISPLAY UNIT

(lb-kg)

301

.01,.02,.05,.1,.2,.5,1,2,5,

10

,20, 50,100

Select display count size of

weigh values.

CAPACITY

(cap)

303

LB

/KG

Display pounds -lb or

kilograms -kg

WM1 ADJUST 1

(wmaI-I)

304

40000

Enter MAXIMUM weight

measurable on scale.

MENU 4. PRESET, BATCHING & ROTATION COUNTER

FEATURES

Does not apply to 2400 indicator

CALIBRATION

DEAD WEIGHT

CAL

(CAL)

802

Calibration method using

weights

SETUP NUMBER

(SETUP)

871

Quick entry method selects

weigh method 1-4lbs, 5-8 kg,

gain 1-9, display counts 1-9

and capacity *1000

CALIBRATION

NUMBER

(CAL)

872

Weight displayed at 0.4mV/V

for these load cells