Flowserve M-series PolyChem User Manual

Page 29

Advertising
background image

USER INSTRUCTIONS PolyChem M-SERIES ENGLISH 71569218 07-11

flowserve.com

Page 29 of 60

®

The coupling guard shown in Figure 5-9 conforms to
the USA standard ASME B15.1, “Safety standard for
mechanical power transmission apparatus.”
Flowserve manufacturing facilities worldwide conform
to local coupling guard regulations.

5.4.2

ClearGuard™ - optional

Flowserve offers as an option a ClearGuard™, which
allows you to see the condition of the coupling (see
Figure 5-10). This guard can be used in place of the
existing clamshell guard described earlier.
Disassembly of the ClearGuard™ is accomplished by
removing the fasteners that hold the two
Clear

Guard™ halves together followed by removing

the foot bolts and rotating the support leg out of the
slot on the guard.

Figure 5-10

Figure 5-11

5.4.3

Trimming instructions

In order to correctly fit the pump/motor configuration,
each guard must be trimmed to a specific length.
This trimming is done on the motor end of the guard.

a)

Measure minimum distance from the center of
mounting hole in the baseplate to the motor.


If clam shell guard proceed to step c

b) Locate a reference center in the slot of the

ClearGuard™ coupling guard flange, see figure 5-
11. Transfer the length measurement to the guard
using this reference center.

c) Trim the motor end of the guard according to the

above measurement. Trimming is best done with a
band saw, but most other types of manual or power
saws give acceptable results. Care must be taken to
ensure that there is no gap larger than 6 mm (0.24
in.) between the motor and the coupling guard.

d)

If motor diameter is smaller than guard

diameter, trim guard so that it extends over the
end of the motor as far as possible.

e)

Deburr the trimmed end with a file or a sharp
knife if ClearGuard™. Care must be taken to
eliminate all sharp edges.


5.4.4

Assembly instructions

Clam shell guard
a)

Mount support leg to each clam shell, figure 5-9.

b)

Attach one half of the guard to the baseplate.

c)

Engage the tabs of guard halves together.

d)

Attach the second support leg to the baseplate.


ClearGuard™
a)

Place the bottom and top halves of the guard
around the coupling.

b)

Install the support legs by inserting and then
rotating the tab on the leg through the slot in the
guard until it comes through and locks the top
and bottom halves of the guard together.

c)

Attach the support legs to the baseplate using
the fasteners and washers provided.

d)

Install fasteners in the holes provided to secure
the guard flanges together.

5.5 Priming and auxiliary supplies


The PolyChem M-series centrifugal pump will not move
liquid unless the pump is primed. A pump is said to be
“primed” when the casing and the suction piping are
completely filled with liquid. Open discharge valves a
slight amount. This will allow any entrapped air to
escape and will normally allow the pump to prime, if the
suction source is above the pump. When a condition
exists where the suction pressure may drop below the
pump’s capability, it is advisable to add a low-pressure
control device to shut the pump down when the
pressure drops below a predetermined minimum.

5.6 Starting the pump


a)

Open the suction valve to full open position. It is
very important to leave the suction valve open
while the pump is operating. Any throttling or
adjusting of flow must be done through the
discharge valve. Partially closing the suction valve
can create serious NPSH and pump performance
problems.

REFERENCE
CENTER

TRIM THIS END

MEASUREMENT STEP a)

Advertising