English, Types, Design – Flowserve OR 42 User Manual

Page 7: Warning, Purpose, Operation, Technical data org spare parts

Advertising
background image

7

Turbidity Detector

Transducer

Sensor

OR 42/3

ORT 4

ORG 12

OR 42/4

ORT 4

ORG 22

OR 52/3

ORT 5

ORG 12

OR 52/4

ORT 5

ORG 22

Types

With three ball valves, three screwed

unions and screwed-in vent nipple

Drying cartridge and turbidity standard

20 ppm (OR 52: also turbidity standard
40 TU/F*) are also supplied

Spare parts kit separately available, see

“Spare Parts”

*) TU/F = turbidity units of Formazin suspension

Design

Light emitter
At a flowrate of 2 l/min and V-shaped flow
through the sensor with a pipe length of 1 m
(DN 10 mm) and four bends: approx. 5 mbar
(

ζ

= 6.1)

Fluids
Water, condensate, drinks etc.

pH value
Up to 10.5 (a pH value of 11 and above will
lead to wear of the glass, depending on the
temperature)

Fluid temperature ranges
0 ... 60 °C with drying cartridge
60 ... 120 °C with vent nipple

Protection
IP 65

Max. ambient temperature
60 °C

Materials

ORG 12

ORG 22

Cover flange

GG-25

1.4580

Wetted parts

0.6025

1.4580

galvanized

Ball valves

Ms 58

1.4436

Screwed unions

St

1.4571

Glass cylinder

Duran 50

Duran 50

O-rings

EPDM

EPDM

Housing

0.6025

0.6025

galvanized galvanized

Cleaning disc

EPDM

EPDM

Light emitter
Glow lamp 12 V/10 W BA 15 s
cable connection via 4 pole connector

Light receiver
Two silicon photo-electric cells,
cable connection via 4 pole connector

Approx. weight
6.8 kg

Dimensions
See Fig. 2

ENGLISH

Use oil and turbidity detector OR 42/OR 52
only for monitoring of transparent liquids to
detect any ingress of insoluble and
refracting foreign matter, for measuring of
turbidity and signal evaluation for indication
and recording as well as for tripping of
alarms, control valves etc.
The equipment must only be installed by
qualified staff.
Before installing the oil and turbidity detector
cut off power supply.
All repairs must be performed only by the
manufacturer. Misuse or any attempted
modification of the equipment will endanger
plant safety.

Important Safety Note

The terminal strip of the ORT4/ORT5 is live
during operation. This presents the danger
of electric shock. Cut off power supply
before opening the equipment or
undertaking any maintenance work

Warning

The oil and turbidity detector OR 42/OR 52
consists of the sensor ORG 12/ORG 22
and the measuring transducer ORT 4/
ORT 5.

Application:

In steam boiler plants: Monitoring of
condensate returned to the boiler for oil
and grease contamination, in particular
for boilers operating without constant
supervision acc. to TRD 604 (2 detectors
are required for 72 h operation)

For raw-water monitoring and water
treatment (sand filters, demineralization
plants, reverse-osmosis plants)

For waste-water monitoring

In breweries and the beverage industry
(filtration, monitoring of wort, quality
assurance etc.)

For salad oil production: Monitoring of
filtering process

For marine applications: Monitoring of

filtering equipment in accordance with
IMO (International Maritime Organization)
resolution MEPC.60(33) – when certain
cold cleaning agents that split up easily
are used which cannot form stable water/
oil emulsions

Purpose

The oil and turbidity detector measures the
intensity of the scattered light produced by
foreign matter that is not dissolved in trans-
parent liquids and exposed to incident light,
with emulsified oils scattering incident light
mainly in the forward direction.
In the oil and turbidity detector the forward
scattered-light principle is applied which
provides a high sensitivity for emulsified
greases and oils and other organic matter.

Operation

Approvals

Feedwater monitoring
TÜV – approval no. 09-96-0132

Feedwater monitoring in marine
applications GL 94855-94 HH
Seeberufsgenossenschaft
GL-OG-94-352002

15 ppm oil detector for bilge monitoring in
engine rooms
GL 85486-94 HH
Seeberufsgenossenschaft
320023

Nominal size
DN 10, connections

3

/

8

"

BSP to DIN ISO 228
Nominal pressure
PN 10
Flowrate
0.5 ... 50 l/min

Technical Data ORG

Spare Parts

The spare parts kit for OR contains:

Designation

Number

Glass cylinder

1

O-ring 25 x 3

2

O-ring 30 x 2

2

O ring 37 x 2

4

Glow lamp

2

For cleaning unit:
Cleaning plunger

1

Ring

1

Cleaning disc

1

Retaining ring 19 x 1.2

1

Fixing disc