Installation, Technical data ort – Flowserve OR 42 User Manual

Page 8

Advertising
background image

8

ORT 4

ORT 5

Wiring

3 cable glands with integral cable clamps PG 11,

3 cable glands with integral cable clamps PG 11,

screw-type terminals with wire guard,

screw-type terminals with wire guard,

max. conductor size 1.5 mm²

max. conductor size 1.5 mm²

Mains supply

230 V ± 10 %, 50/60 Hz

230 V ± 10 %, 50/60 Hz

115 V ± 10 %, 50/60 Hz (option)

115 V ± 10 %, 50/60 Hz (option)

1

24 V ± 10 %, 50/60 Hz (option)

1

24 V ± 10 %, 50/60 Hz (option)

Rated voltage see name plate

Rated voltage see name plate

Fuse

Thermal fuse M 0.1 A

Thermal fuse M 0.2 A

Power consumption

15 VA

25 VA

Measured quantity

Turbidity

Turbidity

Measuring range

0 – 25 ppm

Range I: 0 – 50 TU/F*)

Measuring range see name plate

Range II: 0 – 25 ppm*)
Measuring range see name plate

Lamp voltage for light emitter

No-load voltage 14 V

No-load voltage 14 V

with light emitter connected: < 12 V AC, 10 W

with light emitter connected: < 12 V AC, 10 W

Actual-value output

0 – 1 mA

0 – 20 mA (option 4 – 20 mA)

load 0 ... 120

load 0 ... 500

Output

1 volt-free relay contact

2 volt-free relay contact

Max. contact rating with a switching

Max. contact rating with a switching

voltage of 24/115/230 V AC:

voltage of 24/115/230 V AC:

resistive 4 A, inductive 0.75 A at cos

ϕ

= 0.5

resistive 4 A, inductive 0.75 A at cos

ϕ

= 0.5

Max. contact rating with a switching

Max. contact rating with a switching

voltage of 24 V DC: 4 A

voltage of 24 V DC: 4 A

Life or relay: 10 x 10

6

switching cycles

Life of relay: 10 x 10

6

switching cycles

Contact material: AgCdO 5

µ

hard-gold plated

Contact material: AgCdO 5

µ

hard-gold plated

Switchpoint

Adjustable over complete measuring range

Adjustable over complete measuring range

Delay of response

0 s with external wire linke, 10 s

0.5 – 15 min. adjustable, see name plate

without wire link, see name plate

Indicators and adjustors

1 LED

ALARM

, 1 LED

OPERATION

,

1 LED

ALARM

(

INSTANTANEOUS

), 1 LED

ALARM

(

DELAYED

),

1 LED

MALFUNCTION

, 1 potentiometer for switchpoint,

1 LED

OPERATION

, 1 LED

MALFUNCTION

,

2 potentiometers for adjustment 0/100 %

1 LED analogue display

RESOLUTION

3 %,

1 potentiometer for switchpoint,
2 potentiometers for adjustment 0/100 %,
1 potentiometer for adjustment of delay of response,
1 change-over switch for measuring range I/II,
1 change-over switch for switchpoint/measured value

Protection

IP 54 to DIN 40050

IP 65 to DIN 40050

Max. admissible ambient

0 – 55 °C

0 – 55 °C

temperature

Housing material

Sheet steel, hammer enamel finish

Die-cast aluminium, blue enamel finish

Weight

1.6 kg

5.4 kg

Dimensions

See Fig. 1

See Fig. 1

*) TU/F = turbidity units of Formazin suspension, ppm = parts per million

If the condensate contains suspended
solids (scratching of the glass) installation
as per Fig. 5 should be chosen. If this does
not help and/or a large amount of flash
steam escapes install a separator
upstream of the sensor (Fig. 8)

1. Lid and base can be screwed

turned through 90°.

2. Screw in nipples , screwed unions

and ball valves supplied with the
sensor (use teflon tape or hemp for
sealing).

3. They have to be fitted into the lid and

base or directly at the inlet, outlet or
purging point.

4. The screwed unions facilitate

positioning of the ball valves so that the
levers can be arranged in one plane
(front).

L

M

Screw in vent nipple when the fluid
temperature is above the max. ambient
temperature (Fig. 2).

When the fluid temperature is below the
ambient temperature remove the plastic lid
from the drying cartridge (blue colour)
and screw in the drying cartridge. (Fig. 2)

If the fluid temperature is very high provide
a non-insulated pipe upstream of ORG 12/
ORG 22 which will allow the fluid to cool
down to 120 °C when reaching the sensor.

The second inlet at the lid must be open
for the rinsing water or the calibrating fluid.

In the presence of large amounts of gas
the tapping point must be at the bottom of
the line – use a welding pocket in
accordance with DIN 2618 (Fig. 7, Pos.*)

In the case of very contaminated
condensate: downward flow of the fluid
through the glass cylinder

ORG 12/ORG 22

Install and wire ORG

12/ORG 22 in

accordance with the examples given in
figures 4 to 9.

Mount ORG 12/ORG 22 in a bypass at a
lower level than the main line so that any air
and flash steam can pass through the main
line.

To ensure a flow through the bypass a
throttling point should be provided in the
main line, e. g. GESTRA non-return valve
type RK 66.

Tap product line at the side from the
bottom to prevent air and dirt from
flowing into the bypass.

Any pressure drop upstream of the sensor
which might lead to the release of gases or
to flashing must be avoided.

Installation

F

K

C

E

D

F

K

Technical Data ORT