Ecg/arrhythmia, Adult and pediatric nsr qrs – Fluke Biomedical PS420 User Manual

Page 31

Advertising
background image

Multiparameter Simulator

Simulating Functions

21

ECG/Arrhythmia

The Simulator replicates 37 different types of arrhythmias,
from inconsequential types of PNCs to asystole. In
addition, the Simulator can send waveforms to test any
electrocardiograph, and can accommodate twelve-lead
configurations with independent outputs for each signal
lead referenced to the right leg (RL).

Adult and Pediatric NSR QRS

An adult NSR with a QRS width of 80 ms, or a pediatric
NSR with a QRS width of 40 ms, can be set. These will
remain in effect throughout ECG and arrhythmia
selections until changed by reentering the below codes
and pressing ENTER.

Code Display

Selects:

11

PEDIATRIC

Pediatric NSR with QRS width of
40 ms.

12

ADULT

Adult NSR with QRS width of 80
ms.

NSR

The Simulator replicates fifteen normal sinus rhythms, or
NSRs. These NSRs can be set through direct code entry,
as below, or selected and adjusted through first selecting

9 = NSR. After selecting 9, press ENTER, and then use
the CHANGE keys

to scroll to the available presets.

Code

Display

Selects NSR rate of:

170

NSR 30BPM

30 BPM

171

NSR 40BPM

40 BPM

172

NSR 60BPM

60 BPM

173

NSR 80BPM

80 BPM

174

NSR 100BPM 100 BPM

175

NSR 120BPM 120 BPM

176

NSR 140BPM 140 BPM

177

NSR 160BPM 160 BPM

178

NSR 180BPM 180 BPM

179

NSR 200BPM 200 BPM

180

NSR 220BPM 220 BPM

181

NSR 240BPM 240 BPM

182

NSR 260BPM 260 BPM

183

NSR 280BPM 280 BPM

184

NSR 300BPM 300 BPM

Advertising