Superimposed Artifacts

The Simulator replicates five different artifacts. Their
purpose is to evaluate the effect of these type artifacts on
ECG accuracy. After selecting the artifact, press ENTER
to transmit it to the ECG. The Simulator deactivates the
artifact when you make another ECG or arrhythmia
selection.

Code Display

Selects:

53

50 HZ ARTI

50 Hz artifact (European
lines).

54

60 HZ ARTI

60 Hz artifact (US lines).

55

MUSCLE ART

Muscle artifact.

56

BASE WANDR

Baseline wandering.

57 RESP

ARTIF

Respiration

artifact.

Pacemaker

The Simulator replicates six paced rhythms/signals. After
selecting the required rhythm, press ENTER. Use the
CHANGE keys

to scroll through the available

rhythms.

Code

Display Selects:

58 ASYNCHRONO Asynchronous

pacemaker

rhythm.

59

DEMND FSIN

Pacemaker rhythm with
frequent sinus beats.

60

DEMND OSIN

Pacemaker rhythm with
occasional sinus beats.

61

AV SEQUENT

Atrial ventricular pacemaker
rhythm.

62

NONCAPTURE

Noncapture event. The
Simulator then assumes
asynchronous pacemaker.

63

NON FUNCT

Non-function pacemaker
rhythm.