4 Startup and use

4.1 Startup

4.1.1 Initial check

Connect the sensor to a PoE enabled network, and check that the instrument powers

up properly (see section 2.3). Use a web browser to open the instrument homepage

and check that the serial number of the instrument corresponds to that on the instru-

ment nameplate. In case you have difficulties connecting to the sensor, please see

section 2.3.1.

4.1.2 Calibration check

Wait until normal process conditions occur. The concentration reading is precalibrated

at delivery and a copy of the Sensor calibration certificate is shipped with the sensor.

If the diagnostic message is

NORMAL OPERATION

but the concentration reading does

not agree with the laboratory results, consult Section 5.2, “Calibrating the concentra-

tion measurement”.

4.2 Viewing sensor status

The basic information on the measurements is shown on the main page of the in-

strument (section 2.4.1). More information is shown on the Diagnostics page (sec-

tion 2.4.3).

The measurement result is calculated from the refractive index (n

D

) and process tem-

perature (T) values. Both of these values are available on the main page.

In addition to these measurements, the sensor monitors its internal temperature and

humidity, which both are available on the Diagnostics page. The internal temperature

should not be above 55 °C, and the humidity should be below 30 %.