5 Configuration and calibration

5.1 Configuring the refractometer

In PR-33, all parameter changes are made with a web browser through the parameters

page, see Section 2.4.2.

5.1.1 Signal damping

The system provides the possibility to enter signal damping to diminish the influence of

process noise. The damping is applied to the CONC value (and thus the output signal).

The PR-33 offers two types of signal damping. Exponential (standard) damping works

for most processes and is the standard choice for slow and continuous processes. How-

ever, if the process has fast step changes, linear (fast) damping gives shorter settling

time.

The damping time is set separately. What the damping time means in practice, depends

on the damping type:

In the exponential damping, the damping time is the time it takes for the concentra-

tion measurement to reach half of its final value at a step change. For example, if the

concentration changes from 50 % to 60 % and damping time is 10 s, it takes 10 seconds

for the sensor to display concentration 55 %. A damping time of 5–15 seconds seems

to work best in most cases.

In the linear damping (fast damping), the output is the average of the signal during

damping time. After a step change the signal rises linearly and reaches the final value

after the damping time. Figure 5.1 shows how the damping time affects the measure-

ment.

Note: The factory setting for damping in PR-33-S is 5 s linear. Avoid overdamping, the

signal should not be made insensitive.