Care & maintenance, Troubleshooting – DeVilbiss IFILL PD1000A User Manual

Page 10

Advertising
background image

A-535-CYL

10

E N G L I S H

CARE & MAINTENANCE

The DeVilbiss iFill Oxygen Cylinders should be kept clean and free from moisture and dust. The cylinders should be protected
from extreme temperatures. Clean the cylinders periodically by wiping with a dry, lint-free cloth. Do not clean with a solvent
based cleaning solution. Avoid dropping the cylinder or placing it in a position where it could topple or fall since this can damage
the device. Whenever possible, use a padded carrying bag such as the Hideaway or Walkabout bags to carry the cylinder. This
will help to protect it in case of a fall. Avoid getting fluids or debris such as sand or dirt inside the oxygen cylinder and/or oxygen
connector.

TROUBLESHOOTING

WARNING

Do not attempt to open the device for maintenance or repair. The device contains no

user-serviceable parts. Contact your oxygen provider if service is required.

Oxygen is not being delivered even though
the PulseDose indicator is flashing every
time I breathe.

Use times are different from those stated
in the literature.

PulseDose will not pulse

PulseDose works fine for a couple of min-
utes, then sensitivity seems to drift and
may stop working altogether.

1. Oxygen supply is empty.

1. PulseDose responds to your breath

rate. Your breath rate may vary, which
causes the operation time to vary.

2. Leak in system

1. Cannula is not adjusted properly.

2. Unit is not turned on.

3. Batteries discharged or not installed.
4. Mouth breathing with closed soft

palates.

5. Unit did not reset while changing bat-

teries (red light stays on).

1. Using pediatric cannula or any cannula

that restricts continuous flow capaci-
ty of 10 lpm.

1. Check contents indicator on the

device. If empty, switch cylinders.

1. PulseDose is operating correctly.

2. Contact your Sunrise Medical

provider.

1. Check all cannula connections to

make sure they are tight, and adjust
the cannula to fit comfortably in your
nose. Ensure tubing is not kinked.

2. Turn the rotary selector to the appro-

priate setting.

3. Install new batteries.
4. Breathe through the nose (cannula).

5. Turn the unit off and back on using

the rotary selector.

1. Replace with standard nasal cannula.

Symptoms

Possible Causes

Remedies

C A R E & M A I N T E N A N C E / T R O U B L E S H O O T I N G

Advertising