External power connector, Leds, Ethernet connector leds – Measurement Computing E-1608 User Manual

Page 12: Factory reset button, Screw terminals

Advertising
background image

E-1608 User's Guide

Functional Details

12

The maximum communication distance without using a repeater is 100 meters. You can send your data
100 meters at data speeds of up to 100 Mbps using only one Ethernet cable connected to your computer.

External power connector

Connect the 5 V external power adapter (PS-5V1AEPS) to this connector to provide 5 V external power to the
E-1608.

LEDs

The

Power

LED is steady green when external power between 3.3V to 5.9 V is supplied to the E-1608.

The

Power

LED turns off when:

 power is not supplied by the external supply (make sure that the supply is fully connected to the power

connector)

 the input power is outside of the specified voltage range of the external supply (3.3V to 5.9 V ), causing a

power fault

The E-1608 has an onboard voltage supervisory circuit that monitors the 5 V external power supply.

The

Activity

LED is on when there is a valid host connection, and blinks when a command is received or an

analog input scan is running.

Ethernet connector LEDS

The green

Link/activity

LED on the lower left of the Ethernet connector is on when there is a valid Ethernet

link, and blinks when network activity is detected.

The yellow

Speed

LED on the lower right of the Ethernet connector is on when the transmission speed is

100 Mbps, and off when the transmission speed is 10 Mbps or there is no link.

Factory reset button

Use the factory reset button to reset network configuration settings to the factory default values.

Refer to

Restoring factory default network settings

on page 10 to learn about resetting these values.

Screw terminals

The E-1608 device screw terminals provide the following connections:

 Eight SE or four DIFF (

CH0H/CH0L

to

CH3H/CH3L

) analog input connections

 Eight digital I/O connections (

DIO0

to

DIO7

)

 Two analog output connections (

AOUT0

,

AOUT1

)

 One external clock input (

AICKI

) and one external clock output (

AICKO

) for analog inputs

 One digital trigger input (

TRIG

)

 One counter input (

CTR

)

 One power output (

+VO

)

 Six analog ground (

AGND

) and three digital ground (

GND

) connections

Advertising