Settling time – Measurement Computing PCI-DAS6071 User Manual

Page 37

Advertising
background image

PCI-DAS6071 User's Guide

Specifications

37

Settling Time

Settling time is defined here as the time required for a channel to settle to within a specified accuracy in
response to a full-scale (FS) step. Two channels are scanned at a specified rate. A –FS DC signal is presented to
channel 1; a +FS DC signal is presented to channel 0.

Table 6. Settling time specifications

Condition

Range

Accuracy

±0.012%
(±0.5 LSB)

±0.024%
(±1.0 LSB)

±0.098%
(±4.0 LSB)

Same range to
same range

±10 V

Typ
Max

3.0 µS
5.0 µS

1.9 µS
2.5 µS

1.9 µS
2.0 µS

±5 V

Typ
Max

3.0 µS
5.0 µS

1.9 µS
2.5 µS

1.2 µS
1.5 µS

±2.5 V

Typ
Max

3.0 µS
5.0 µS

1.9 µS
2.5 µS

1.2 µS
1.3 µS

±1 V

Typ
Max

3.0 µS
5.0 µS

1.9 µS
2.5 µS

1.2 µS
1.3 µS

±500 mV

Typ
Max

3.0 µS
5.0 µS

1.9 µS
2.5 µS

1.2 µS
1.3 µS

±250 mV

Typ
Max

3.0 µS
5.0 µS

1.9 µS
2.5 µS

1.2 µS
1.3 µS

±100 mV

Typ
Max

3.0 µS
5.0 µS

1.9 µS
2.5 µS

1.2 µS
1.3 µS

±50 mV

Typ
Max

3.0 µS
5.0 µS

1.9 µS
2.5 µS

1.2 µS
1.5 µS

0 to 10 V

Typ
Max

3.0 µS
5.0 µS

1.9 µS
2.5 µS

1.2 µS
1.5 µS

0 to 5 V

Typ
Max

3.0 µS
5.0 µS

1.9 µS
2.5 µS

1.2 µS
1.3 µS

0 to 2 V

Typ
Max

3.0 µS
5.0 µS

1.9 µS
2.5 µS

1.2 µS
1.3 µS

0 to 1 V

Typ
Max

3.0 µS
5.0 µS

1.9 µS
2.5 µS

1.2 µS
1.3 µS

0 to 500 mV

Typ
Max

3.0 µS
5.0 µS

1.9 µS
2.5 µS

1.2 µS
1.3 µS

0 to 200 mV

Typ
Max

3.0 µS
5.0 µS

1.9 µS
2.5 µS

1.2 µS
1.3 µS

0 to 100 mV

Typ
Max

3.0 µS
5.0 µS

1.9 µS
2.5 µS

1.2 µS
1.5 µS

Advertising