Four-wire, single-sensor, Semiconductor sensor measurements, Figure 7 – Measurement Computing USB-TEMP-AI User Manual

Page 15: Figure 8

Advertising
background image

USB-TEMP-AI User's Guide

Signal I/O Connections

15

Four-wire, single-sensor

A four-wire, single-sensor connected to the first channel of a channel pair is shown in Figure 7.

Figure 7. Four-wire, single RTD or thermistor sensor measurement configuration

A four-wire, single-sensor connected to the second channel of a channel pair is shown in Figure 8.

Figure 8. Four-wire, single RTD or thermistor sensor measurement configuration

A four-wire, two-sensor measurement configuration is shown in Figure 9.

Figure 9. Four-wire, two RTD or thermistor sensors measurement configuration

When configured for four-wire, two sensor mode, both sensors must be connected to obtain proper
measurements.

Semiconductor sensor measurements

Semiconductor sensors are suitable over a range of approximately –40 C to 125 C, where an accuracy of
±2 C is adequate. The temperature measurement range of a semiconductor sensor is small when compared to
thermocouples and RTDs. However, semiconductor sensors can be accurate, inexpensive and easy to interface
with other electronics for display and control.

The USB-TEMP-AI makes high-resolution measurements of semiconductor sensors, and returns a 32-bit
floating point value in either voltage or temperature.

Use InstaCal to select the sensor type (LM35, TMP35 or equivalent) and the sensor input channel to connect the
sensor.