Calibration

9

Overview …… 9-1

Using DaqCal.exe …… 9-1

Overview

The DaqBook/2000 and DaqOEM/2000 Series devices are digitally calibrated at the factory. The digital

method involves storing a correction factor on the circuit board for each range at the time of calibration.

Whenever a particular range is selected, the appropriate calibration constant is automatically applied to a

compensating DAC, thereby calibrating the specific range. The result is that readings generated by the A/D

are already calibrated, and do not require additional processing.
This is significantly better than the calibration method in which readings are adjusted in software after they

are transferred to the host PC. That method reduces the dynamic range of the A/D, and can adversely

affect the speed at which the PC obtains a calibrated reading.
The DaqBook/2000 and DaqOEM/2000 Series devices include a user-cal mode, whereby the user can

adjust the calibration of the unit’s board, without destroying the supplied factory calibration. This is

accomplished with two distinct calibration tables in the unit’s on-board EPROM, one which contains the

factory calibration factors, and the other which is available for user calibration.

Using DaqCal.exe

When calibration adjustments are needed they should be completed in the following order:

1.

2.

3.

4.

5.

6.

7.

8.

PGA Input and Output Offset

Sample/Hold Offset

A/D Offset and Gain

VDC Voltage Reference

DAC0 Full-Scale

1

DAC1 Full-Scale

1

DAC2 Full-Scale

1

DAC3 Full-Scale

1

1

DAC Full-Scale only applies to DaqBook/2020, DaBook/2001, and DaqOEM/2001.


A Windows-based program, DaqCal.exe, is used to calibrate Daq systems, including analog expansion

cards. DaqCal.exe is used in conjunction with:

a 4.5 digit, digital multi-meter

an adjustable voltage calibrator

an ambient temperature meter


To use the calibration program:

1. Launch DaqCal.

Note: DaqCal is installed automatically from your data acquisition CD as a part of product

support. This takes place during software installation. DaqCal’s default location is the

IOtech DaqX Software folder, in the Programs group.

2. When DaqCal opens you will be prompted to select your device from a list. After doing so,

simply follow the illustrated on-screen instructions.