Using accelerometers – Measurement Computing ZonicBook 618E rev.3.4 User Manual

Page 61

Advertising
background image

Using Accelerometers

Overview

A low-impedance piezoelectric accelerometer consists of a piezoelectric crystal and an electronic amplifier.
When stretched or compressed, the two crystal surfaces develop a charge variation that is related to the
amount of stress, shock, or vibration on the crystal. The amplifier outputs a corresponding signal and
transforms the sensor’s high impedance to a lower output impedance of a few hundred ohms. Note that, in
addition to acceleration, these sensors can also measure pressure and force.

The circuit requires only two wires (coax or twisted pair) to transmit both power and signal. At low
impedance, the system is insensitive to external or “triboelectric” cable noise. Cable length does not affect
sensitivity.

The following figure shows a simple sensor-WBK18 connection. The voltage developed across R is
applied to the gate of the MOSFET. The MOSFET is powered from a constant current source of
4 mA and 30 volts.

Using a Sensor with a WBK18

This diagram also applies to ZonicBook/618E.


The MOSFET circuit will bias at approximately 12 V in the quiet state. As the system is excited, voltage is
developed across the crystal and applied to the gate of the MOSFET. This voltage will cause linear
variation in the impedance of the MOSFET and a proportional change in bias voltage. This voltage change
will be coupled to the WBK18 input amplifier through the capacitor C. The value of R and the internal
capacitance of the piezoelectric crystal control the low frequency corner. Units weighing only a few grams
can provide high-level outputs up to 1 V/g with response to frequencies below 1 Hz.

Accelerometer Specification Parameters

Noise in Accelerometers

The noise floor or resolution specifies the lowest discernible amplitude (minimum “g”) that can be
measured. There are two main sources of noise as follows:

Noise from the crystal and microcircuit inside the accelerometer. Some types of crystals, such as

quartz, are inherently noisier than others. A good noise floor is 10 to 20 µV.

Noise from electrical activity on the mounting surface. Since the signal from the accelerometer is a

voltage, 60 Hz or other voltages (ground loop, etc) can interfere with the signal. The best protection
is to electrically isolate the accelerometer.

ZonicBook/618E

878595

Analog Signals 6-13

Advertising
This manual is related to the following products: