Current Requirements (in Amps)

Products

DBK30A

14 VDC

DBK30A

28 VDC

DBK34A

12 VDC

DBK34A

24 VDC

TR-40U

15 VDC

ZonicBook/618E (alone)

2.5 1.3 2.8 1.4 2.3

WBK18

1.4 0.75 1.6 0.80 1.3

WBK30

0.01 0.005 0.01 0.006 0.01

Worksheet for Power Requirements

Product

Qty

×

Amps

=

Totals

WaveBook/516E

WBK18

× =

WBK30

× =

Maximum

Amps

ZonicBook/618E Product Power Supplies

Item

Name/Description

Capacity

TR-40U

AC Power Adapter (shipped with ZonicBook/618E &

WBK18)

90-264 VAC input;

3.33 A @ 15 VDC

DBK30A

Rechargeable Battery/Excitation Module (optional)

12-14 VDC, or 24-28

VDC

3.4 A-hr @ 14 VDC

DBK34A

UPS (uninterruptible Power Supply)/Battery Module

(optional)

12 VDC, or 24 VDC

5.0 A-hr @ 12 VDC

Other

10 to 30 VCD source, such as a vehicle battery via a

CA-116 cable.

Depends on source

Input voltage to the ZonicBook/618E and to the WBK18 expansion modules must be in the range of
10 to 30 VDC and can come from an AC-to-DC adapter or from another source, such as a battery.

Available AC-to-DC adapters include the TR-40U, which has an input of 90 to 264 VAC and an output
of 3.33 amps @ 15 VDC.

Battery options include the DBK30A, DBK34A, and other 10 to 30 VDC sources, such as car batteries.
The DBK30A provides 14 VDC and when fully-charged has a storage capacity of 3.4 A

⋅hr; car batteries

have much higher capacities. The basic formula for battery life is:

Runtime (hr) = Battery capacity (Ahr) / Current load (A)

Battery life and performance depend on various factors including battery type, condition,
charge level, and ambient temperature. Be sure you consider these factors, especially
when runtime is a critical.

Reference Note:
For information regarding alternate power supplies, refer to Appendix C, Power Options.