RLE GD200 User Manual

Option accessories

Advertising
background image

D

C

B

A

A

B

C

D

1

2

3

4

5

6

7

8

8

7

6

5

4

3

2

1

THE INFORMATION CONTAINED IN THIS
DRAWING IS THE SOLE PROPERTY OF
Quatrosense Environmental Ltd. ANY
REPRODUCTION IN PART OR AS A WHOLE
WITHOUT THE WRITTEN PERMISSION OF
Quatrosense Environmental Ltd. IS
PROHIBITED.

PROPRIETARY AND CONFIDENTIAL

NEXT ASSY

USED ON

APPLICATION

DIMENSIONS ARE IN INCHES
TOLERANCES:
FRACTIONAL 1/32
ANGULAR:
MACH .5 degrees BEND
TWO PLACE DECIMAL .02
THREE PLACE DECIMAL .010

INTERPRET GEOMETRIC
TOLERANCING PER:

MATERIAL

FINISH

DRAWN

CHECKED

ENG APPR.

MFG APPR.

Q.A.

COMMENTS:

DATE

NAME

Quatrosense Environmental Ltd

TITLE:

SIZE

B

DWG. NO.

REV

WEIGHT:

SCALE: 1:8

UNLESS OTHERWISE SPECIFIED:

B

XY

XY

XY

2010/02/25

2010/02/25

2010/02/25

Q5

INSTALLATION DRAWING

SHEET 1 OF 5

85950-002-000

DO NOT SCALE DRAWING

SPECIFICATION

INPUT POWER:
+24VDC nominal, range: 18 to 30VDC 0.3A DC Total Max.
~24VAC nominal, range: 15 to 24VAC 50/60HZ 0.3A AC Total Max.

(AC must not be grounded)

FUSE:

F2 on Main Board: Polyswitch 750mA
Polyswitch device resets after the fault is cleared and power to the
circuit is removed

SENSOR:

Combustible gases: Catalytic
Toxic gases and Oxygen: Electrochemical

OUTPUT SIGNAL:

RS-485 with OPTIMUX PROTOCOL AND MODBUS PROTOCOL
AVAILABLE CONTROLLER: M-CONTROLLER or Q4 CONTROLLER
4-20mA Analog Output, 1-5VDC, 2-10VDC Output
3X SPDT RELAYS: 1.0A MAX. @30VDC (RESISTIVE LOAD)
0.3A MAX. @125VAC (RESISTIVE LOAD)

ENCLOSURE:

IP 66 & NEMA 4, 4X, 12 & 13

OPERATING TEMPERATURE:

-40°C to 70°C, depends on sensor specification

AMBIENT HUMIDITY:

5% TO 95% RH (NON- CONDENSING)

STORAGE TEMPERATURE:

0°C to 20°C, depends on sensor specification

SIZE:

150mm X 90mm X 65mm

WEIGHT:

LESS THAN 0.5lbs

Option Accessories:

Pump-thru & Cal Cap Kit
SKU#: 85930-006-000

Splash Guard Kit
SKU#: 85930-007-000

Duct Mount Adapter
Kit: 85930-040-000

Q-View & USB-RS485 Converter Kit: 85930-004-000

*Option Accessories are not included in Q5 Standard Package.

REVISIONS

ECN

REV.

DESCRIPTION

DATE

DRAW

CHECK APPROVED

924

A

Initial Release

2010/02/25

XY

XY

XY

929

B

Update Duct Mount Kit

2010/08/23

XY

XY

XY