7 Environmental Conditions

7.1 The IRT 1557 has been designed to perform tests

and measurements in a dry environment.


7.2 Operating Altitude 0 to 2000 metres

7.3 Overvoltage category IEC 60664/IEC 61010,

300V Category IV,
600V Category III.
1000V Category II.



7.4 Pollution degree 2 according to IEC 61010-1.

7.5 Protective system IP40 according to IEC 60529.

7.6 Electromagnetic compatibility (EMC). Interference

immunity and emitted interference conforming to
IEC 61326-1.


7.7 Operating temperature range of 0°C to 40°C,

without moisture condensation.


7.8 The IRT 1557 can be stored at any temperature in

the range -25°C to +65°C (relative humidity up to
90%). The batteries should be taken out of the
instrument for storage.