SL Power Electronics GPM140 User Manual

Gpm140 series installation instructions

Advertising
background image

CONDOR DC POWER SUPPLIES INC.

2311 STATHAM PKWY

OXNARD, CA 93033 + 805-486-4565

RATINGS:

Input: 100-240 V ac, 4.0 A, 50/60 Hz

Output:

MAXIMUM OUTPUT AMPS AND WATTS

CONVECTION COOLING

FORCED AIR COOLING (1)

Model

Volts

Without Cover

With Cover

With & Without Cover

Isc

GPM140-5

5 26.0 A 130 W

24.0 A 120 W

30.0 A 160 W

28.5 A

GPM140-12

12 11.7 A 140 W

10.0 A 120 W

13.3 A 160 W

18.2 A

GPM140-13

13 10.8 A 140 W

9.20 A 120 W

12.3 A 160 W

16.6 A

GPM140-15

15

9.3 A 140 W

8.0 A 120 W

10.7 A 160 W

16.6 A

GPM140-24

24

5.8 A 140 W

5.8 A 140 W

6.7 A 160 W

13.2 A

GPM140-28

28

5.0 A 140 W

5.0 A 140 W

5.7 A 160 W

12.5 A

GPM140-30

30

4.7 A 140 W

4.7 A 140 W

5.0 A 160 W

12.0 A

Notes: 1. Minimum airflow for forced air cooling is 26 cfm.

2. Maximum ambient temperature for continuous output power specified above is 50 °C.

3. Isc = Maximum output short circuit current.

4. Maximum Relative Humidity 96 %, no condensation.

5. Storage: -40 to +85 °C. Units should be allowed to warm-up under non-condensing conditions before

application of power.

SAFETY DECLARATION: Condor DC Power Supplies, Inc. declares under our sole responsibility that all

models listed above are in conformity with the applicable requirements of EN 60950 following the provisions

of the Low Voltage Directive 73/23/EEC.

CERTIFICATION: All models are Certified to be in compliance with the applicable requirements of UL 2601-1, CSA 22.2

No. 234 (Level 3) (with additional tests to C22.2 No 601.1-M90 per T.I.L. CA-08), and EN 60601-1: 1990 +A13.

CLASSIFICATION:

(5.1) Protection against electric shock = Class I

(In accordance with sub-

(5.2) Degree of protection against electric shock = Signal output or intermediate

clause 5 of EN 60601-1)

(5.3) Protection against harmful ingress of water = Ordinary (no protection)

(5.5) Has not been evaluated for use in the presence of a flammable anaesthetic mixture with

air, oxygen, or nitrous oxide. This evaluation is made on the end equipment by the OEM.

(5.6) Mode of operation = Continuous

GROUNDING: The Protective Earth (ground) terminal J1, pin 5 (TB1-3) must be bonded to Protective Earth in the host

equipment. Using the Protective Earth terminal on the supply for grounding the host equipment is not recommended. A

separate dedicated grounding point should be used.
OUTPUTS: The output common should be connected to Protective Earth in the end application. The output is intended for

Protectively Earthed Signal Output and Intermediate Circuits only. The output is not acceptable for patient connection

without additional isolation. All DC outputs are SELV under normal and single fault conditions.
OVERVOLTAGE PROTECTION: The output is monitored for an overvoltage condition. The trip-point for a 5 volt

output is 5.6 to 6.8 volts. In some applications where an overvoltage condition could result in a hazard as defined in

applicable safety standards, redundant or additional overvoltage protection may be required. Consult factory for details.

CAUTION: When performing Dielectric Strength Tests, catastrophic failure of the unit may result if a Dielectric Strength

test voltage greater than 1800 V ac is applied between primary and secondary circuits. The components providing isolation

from primary to secondary cannot be tested while installed in the power supply without overstressing basic (primary to

ground) insulation. All isolating components are individually 100 % tested at 4800 V ac prior to installation.

41-33340-0001 Rev. E 09/02/04